Assessment of a new cell culture perfusion apparatus for in vitro chronic toxicity testing. Part 1: Technical description
Main Article Content
Abstract
In vitro models for chronic toxicity, defined as a recurring exposure to compounds over a prolonged period of time, are still underrepresented in drug evaluation processes. The classical approach to cell culture is not readily suitable to long term repetitive applications. Therefore, we assessed the use of a commercially available perfusion cell culture apparatus in its applicability to chronic renal toxicity testing and describe the technical aspects of adopting the perfusion cell culture system to our purposes. It was apparent that there is a subtle dynamic difference between human renal proximal tubular cells cultured under perfusion and static conditions as illustrated by the accumulation of lactate dehydrogenase (LDH) and the secondary metabolism of resazurin to hydroresorufin, which occurred only under static conditions. The major achievement was the standardisation of the handling of this system with regard to cell cultivation, pH regulation, temperature regulation, and reproducibility of common toxicity endpoints.
Article Details
Articles are distributed under the terms of the Creative Commons Attribution 4.0 International license (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium, provided the original work is appropriately cited (CC-BY). Copyright on any article in ALTEX is retained by the author(s).